Nexus replaces the proprietary spectrometer of an MRI system with two commercial PCIe measurement cards from Spectrum Instrumentation, an arbitrary waveform generator (transmit) and a digitizer (receive), operated by open-source Python software. The console is designed for pulse sequences that are defined in the open Pulseq format. Using PyPulseq, they can be built directly within the Nexus framework or loaded from a valid .seq file. From the Pulseq sequence, the corresponding RF and gradient waveforms are calculated and streamed to the transmit card without the need for a vendor-specific sequence compiler. Received signals are demodulated, phase-corrected and decimated on the host and exported in the ISMRMRD format for reconstruction with any compatible framework. The console was developed and validated on a ~50 mT low-field MRI system presented here. It is not tied to a specific field strength or hardware setup: amplifiers, coils and system limits are described in a configuration file.
Nexus Console or Nexus CLI?
This documentation is aimed at developers and researchers who want to build on the console software itself, e.g. to develop new sequences, acquisition workflows or hardware integrations. Using the nexus-console repository alone requires a custom set of experiments or a runner script that handles all tasks around the acquisition, such as loading a sequence, setting the acquisition parameter and saving the data.
If you only want to execute sequences and experiments, use the Nexus CLI instead. It runs Pulseq .seq files and protocols, exports the data as ISMRMRD, manages the acquisition parameter and provides basic calibrations, without writing Python.
-
Setup
Connect the measurement cards, install the software and the acquisition service, and run a first acquisition.
-
System guide
How the console works: hardware, configuration, sequence calculation, acquisition and postprocessing.
-
Examples
Example experiments, from spin-echo spectroscopy and T2 relaxometry to 3D imaging and remote acquisition.
-
API reference
Classes and functions of the console package, generated from the docstrings of the source code.
-
Nexus CLI
Command-line client to run Pulseq sequences, edit the acquisition parameter and calibrate the system.
-
Issues and discussions
Facing a problem? Check the existing issues and discussions, or open a new one if it is not covered.
Citation
Tip
If Nexus contributes to your research, please cite our paper:
Schote D, Silemek B, O'Reilly T, Seifert F, Assmy JL, Kolbitsch C, Webb AG, Winter L. Nexus: A versatile console for advanced low-field MRI. Magn Reson Med. 2025;93(5):2224–2238. doi:10.1002/mrm.30406
BibTeX
@article{schote2025nexus,
title = {Nexus: A versatile console for advanced low-field {MRI}},
author = {Schote, David and Silemek, Berk and O'Reilly, Thomas and Seifert, Frank and
Assmy, Jan-Lukas and Kolbitsch, Christoph and Webb, Andrew G. and Winter, Lukas},
journal = {Magnetic Resonance in Medicine},
volume = {93},
number = {5},
pages = {2224--2238},
year = {2025},
doi = {10.1002/mrm.30406}
}
Acknowledgments
This work is part of the Metrology for Artificial Intelligence for Medicine (M4AIM) project, funded by the Federal Ministry of Economic Affairs and Climate Action (BMWK) as part of the QI-Digital initiative. The projects 21NRM05 STASIS and 22HLT02 A4IM have received funding from the European Partnership on Metrology, cofinanced by the European Union's Horizon Europe Research and Innovation Program and by the Participating States. Partial support is provided by a European Research Council Advanced Grant (PASMAR 101021218). This work is supported by the Open Source Imaging Initiative (OSI²).