Postprocessing

Each ADC gate recorded by the digitizer is converted from raw ADC values into a complex baseband signal at the dwell time of the PyPulseq ADC event. The processing runs in parallel to the acquisition: the RX thread submits every gate as an RxData object to the RxProcessor, a pool of num_processing_workers worker processes that is started once when AcquisitionControl is created. The raw samples are passed to the workers via shared memory. With num_processing_workers: 0, all gates are processed in the main process after the acquisition. After processing, the raw samples are discarded unless run(store_unprocessed=True) is used.

%%{init: {"fontFamily": "Inter, sans-serif"}}%%
flowchart LR
    G["Raw data<br/>(int16)"] --> S["Scaling<br/>to mV"]
    S --> D["Demodulation"]
    D --> C["Phase<br/>correction"]
    C --> Dec["Decimation"]
    Dec --> O["Processed data<br/>(complex float)"]

    classDef sw fill:#2f6bff22,stroke:#2f6bff,stroke-width:1.5px
    class G,S,D,C,Dec,O sw

Processing steps

The steps are implemented in RxData.process_data and applied to each receive channel of a gate:

Step Operation
Scaling The int16 values are converted to mV with the input range of the channel: value × channel_max_amplitude / 32768.
Demodulation The signal is multiplied by \(e^{-j 2\pi f_\text{d} t}\), with the demodulation frequency \(f_\text{d}\) = Larmor frequency + frequency offset of the ADC event and the time \(t\) from the start of the gate.
Phase correction The phase measured with the phase reference (see below) is removed, and the phase offset of the ADC event is applied, i.e. the signal is multiplied by \(e^{j \varphi_\text{ADC}}\).
Decimation The signal is low-pass filtered and down-sampled to the dwell time of the ADC event (see Decimation filters). The decimation factor is the ratio of the ADC dwell time and the sampling interval of the digitizer (1/RxConfiguration.sampling_rate). The result is multiplied by 2, which compensates the signal component at \(-2 f_\text{d}\) that is removed by the low-pass filter.

The processed data of a gate contain num_samples + 2 × num_samples_discard complex samples per receive channel, i.e. the ADC samples plus the samples of the gate extension on both sides (see ADC events).

Phase reference

The gate detection of the digitizer is not phase-locked to the replay of the AWG, so the phase of the demodulated signal would vary from gate to gate. To correct this, the digitizer samples the phase reference, a 1.095 MHz square wave that the AWG outputs at the start of each gate (see Sequence calculation). The phase of the reference is determined by demodulating the first ≤ 1000 reference samples at \(f_\text{ref}\) = 1.095 MHz and summing them. The phase at \(f_\text{ref}\) corresponds to a time offset, which is converted to the phase at the demodulation frequency:

\[ \varphi_\text{ref} = \arg\left( \sum r(t)\, e^{-j 2\pi f_\text{ref} t} \right) \cdot \frac{f_\text{d}}{f_\text{ref}} \]

The signal of the gate is then multiplied by \(e^{-j \varphi_\text{ref}}\).

Decimation filters

The filter is selected by AcquisitionParameter.ddc_method (DDCMethod). The decimation factor must be greater than 1.

Method Implementation
FIR (default) scipy.signal.decimate with an FIR anti-aliasing filter
CIC Five-stage cascaded integrator-comb filter [5] that decimates by about half the decimation factor, followed by an FIR decimation by 2
AVG Strided convolution with a windowed kernel that spans 8 decimated samples

Data structures

Object Content
RxData One ADC event of one average: metadata of the event, PyPulseq labels, time stamp of the gate, processed_data (complex, shape (channels, samples)) and optionally raw_data
AcquisitionData Result of AcquisitionControl.run: flat list of all RxData, ordered by average and then by ADC event, the sequence, the acquisition parameters and metadata including the device configuration

Storage

All files of an acquisition are written to the folder <session>/<acquisition id>/. The session folder is <home>/nexus-console/<YYYY-MM-DD>-session by default; the acquisition id consists of date, time and sequence name, e.g. 2026-10-29-143012-se_spectrum.

Method Output
AcquisitionData.save() rx_data.h5: metadata and one group per RxData with its metadata, processed_data and, if kept, raw_data
meta.json: software version, date and time, acquisition id, acquisition parameters, sequence name, duration and definitions, device configuration and additional information from add_info
sequence.seq: the PyPulseq sequence
<key>.npy: additional arrays added with add_data
AcquisitionData.save_ismrmrd() data.mrd in ISMRMRD format [4], see below

An existing meta.json or sequence.seq is only overwritten with save(overwrite=True).

ISMRMRD export

AcquisitionData.save_ismrmrd(header) writes the processed data of all ADC events to data.mrd which is the open source ISMRMRD format, so that they can be reconstructed with external frameworks such as MRpro.

The header argument provides the encoding information required for the reconstruction, e.g. matrix size, field of view and encoding limits:

header Used header
ismrmrd.xsd.ismrmrdHeader The given header, e.g. created together with the sequence
Path to an MRD file The XML header of that file
None (default) An empty header without encoding information

The following entries of the header are set by the console and overwrite existing values:

Header entry Value
measurementInformation Acquisition id as measurement id, series date and time
experimentalConditions Larmor frequency in Hz
acquisitionSystemInformation Number of receive channels, vendor osi2, model Nexus_<version>, field strength that corresponds to the Larmor frequency

Acquisitions

Each RxData with processed data becomes one ISMRMRD acquisition, in the order of AcquisitionData.receive_data:

Acquisition field Value
data processed_data of the ADC event
scan_counter Position in receive_data
number_of_samples, active_channels Samples (including the gate extension) and receive channels of the ADC event
discard_pre, discard_post num_samples_discard, i.e. the samples of the gate extension
center_sample num_samples / 2, rounded down
sample_time_us Dwell time of the ADC event in µs
acquisition_time_stamp Start of the gate in µs, relative to the start of the digitizer
idx counters and flags From the PyPulseq labels of the ADC event (see below)
read_dir, phase_dir, slice_dir From the channel assignment (see below)
traj k-space trajectory from the sequence (see below)

The PyPulseq labels set since the previous ADC event are stored in the RxData object and mapped to the counters and flags of the acquisition. Since averaging can be configured via the acquisition parameter or via Pulseq, the RxData attribute is used, which already considers both options.

Pulseq label stored in RxData ISMRMRD field
LIN kspace_encode_step_1
PAR kspace_encode_step_2
SLC slice
RxData.average_index (AVG) average
ECO contrast
PHS phase
REP repetition
SET set
SEG segment
NOISE Flag ACQ_IS_NOISE_MEASUREMENT
REF Flag ACQ_IS_PARALLEL_CALIBRATION
IMA Flag ACQ_IS_PARALLEL_CALIBRATION_AND_IMAGING
REV Flag ACQ_IS_REVERSE
NAV Flag ACQ_IS_NAVIGATION_DATA

Orientation

The directions are given in the patient coordinate system (LPS), assuming a patient in supine, head-first position and the following orientation of the physical gradient channels:

Physical channel Patient direction LPS vector
1 Inferior → superior (0, 0, 1)
2 Posterior → anterior (0, −1, 0)
3 Right → left (1, 0, 0)

read_dir, phase_dir and slice_dir are the vectors of the physical channels assigned to the logical axes x, y and z by channel_assignment. They are identical for all acquisitions.

Trajectory

The k-space trajectory of all ADC samples is calculated from the sequence with PyPulseq (calculate_kspace). The resulting array always has the shape (3, num_ro_samples), where the first dimension accounts for the three spatial dimensions. The actual trajectory dimension is determined by the number of axes containing non-zero coordinates. If a trajectory exists, i.e. the number of non-zero axes is greater than zero, the relevant trajectory section is written to the corresponding ISMRMRD acquisition object. Based on the ADC (RxData) index, the trajectory segment relevant to the ADC event is extracted and inserted into the ISMRMRD acquisition. For the pre- and post-samples of the ADC gate, which are defined by the ADC dead time, the trajectory is padded with the first and last coordinate of the ADC trajectory, respectively.