Device configuration
The device configuration describes the hardware of the MRI system: the measurement cards, the amplifiers, the gradient coils and the system limits. It is read from a YAML file when AcquisitionControl is created and remains fixed for the session. Parameters that change between measurements, such as the Larmor frequency, are passed with every sequence as AcquisitionParameter.
flowchart LR
Y(["YAML file"]) -->|"load_nexus_config"| N["NexusConfiguration"]
N -->|"TxConfiguration"| T["TxCard"]
N -->|"RxConfiguration"| R["RxCard"]
N -->|"RxConfiguration"| P["RxProcessor"]
N -->|"TxConfiguration,<br/>SystemLimits"| S["SequenceProvider"]
N -->|"complete copy"| D["AcquisitionData"]
classDef sw fill:#2f6bff22,stroke:#2f6bff,stroke-width:1.5px
classDef iface fill:#8e44ad22,stroke:#8e44ad,stroke-width:1.5px
class N,T,R,P,S,D sw
class Y iface
The file is validated by NexusConfiguration and consists of the three sections described below. Unknown keys are ignored. A copy of the complete configuration is stored with every acquisition in meta["device_configuration"].
TxConfiguration
Transmit card (AWG), validated by TxConfiguration. channel_max_amplitude and channel_filter_type have four entries in the order [RF, gradient 1, gradient 2, gradient 3]; gradient_efficiency and gpa_gain have three entries, one per physical gradient channel.
| Attribute | Unit | Description |
|---|---|---|
device_path |
– | Device path of the transmit card, e.g. /dev/spcm1 |
sampling_rate |
MHz | Sampling rate of the transmit card as integer; may differ from RxConfiguration.sampling_rate (see note below) |
channel_max_amplitude |
mV | Output range per channel, 1–6000 mV into 50 Ω |
channel_filter_type |
– | Output filter setting per channel, 0–3 (see card manual) |
rf_terminated_50ohm |
– | true if the RF output is terminated into 50 Ω, false for high impedance (the output voltage doubles) |
gradients_terminated_50ohm |
– | true if the gradient outputs are terminated into 50 Ω, false for high impedance (the output voltage doubles) |
gradient_efficiency |
T/m/A | Gradient coil efficiency per physical gradient channel |
gpa_gain |
A/V | Transconductance of the gradient power amplifier per physical gradient channel |
rf_to_mvolt |
mV/Hz | Conversion of the PyPulseq RF amplitude to the output voltage |
rf_gain_lut_path |
– | Optional, default none. Path to a lookup table for the linearization of the RF power amplifier (.npy, 65536 int16 values) |
RxConfiguration
Receive card (digitizer), validated by RxConfiguration. The parameters channel_enable, channel_max_amplitude and channel_terminated_50ohm are defined by a list containing one entry per physical channel (i.e. the length of the list should correspond to max_available_channels).
| Attribute | Unit | Description |
|---|---|---|
device_path |
– | Device path of the receive card, e.g. /dev/spcm0 |
sampling_rate |
MHz | Sampling rate of the receive card as integer; may differ from TxConfiguration.sampling_rate (see note below) |
max_available_channels |
– | Number of input channels of the card and length of all channel lists (8 for the M2p.5933-x4) |
channel_enable |
– | Enable flag per channel; 1, 2, 4 or 8 channels must be enabled |
channel_max_amplitude |
mV | Input range per channel: 200, 500, 1000, 2000, 5000 or 10000 |
channel_terminated_50ohm |
– | Input termination per channel, true for 50 Ω, false for high impedance |
num_processing_workers |
– | Optional, default 1. Number of worker processes for the parallel processing of the receive data; 0 processes the data in the main process after the acquisition |
Sampling rates of transmit and receive card
Both cards share the clock of the receive card, but each card sets its own sampling rate. The hardware therefore allows different sampling rates, e.g. 20 MHz for the receive card and 10 MHz for the transmit card. The current software version does not support this yet: the sequence calculation and the processing of the receive data assume a common sampling rate, so NexusConfiguration rejects a configuration in which TxConfiguration.sampling_rate and RxConfiguration.sampling_rate differ. Support for different sampling rates will be added in a future update.
SystemLimits
Limits of the MR system, validated by SystemLimits and passed to PyPulseq as Opts (AcquisitionControl.get_sequence_system()). max_grad, max_slew, block_duration_raster, grad_raster_time and adc_raster_time must be positive, all other values must not be negative. Numbers in exponent notation need a decimal point, e.g. 20.e-6: YAML reads 20e-6 as a string, which the model rejects.
Limitations of the system limits
When a Pulseq sequence is read from a file, it is loaded with this configuration. When a Sequence object is passed, its Opts object is likewise overwritten by this configuration. This may cause the sequence timing check to fail if the block durations stored in the sequence are incompatible with the raster times defined here. Note also that max_grad and max_slew are only used when building a sequence. A sequence passed as an existing Sequence object or read from a file is not checked for its maximum gradient amplitude or slew rate. An error is only raised during the calculation of the sequence waveforms if the maximum amplitude of a channel is exceeded.
| Attribute | Unit | Description |
|---|---|---|
max_grad |
Hz/m | Maximum gradient amplitude |
max_slew |
Hz/m/s | Maximum gradient slew rate |
rf_dead_time |
s | Dead time before an RF pulse; the RF unblanking signal is set this time before the pulse |
rf_ringdown_time |
s | Ring-down time after an RF pulse |
adc_dead_time |
s | Dead time around an ADC event; the ADC gate is extended by this time on both sides |
block_duration_raster |
s | Raster time of block durations |
rf_raster_time |
s | Raster time of RF waveforms |
grad_raster_time |
s | Raster time of gradient waveforms |
adc_raster_time |
s | Raster time of ADC dwell times |
B0 |
T | Optional, default 0.05. Static magnetic field strength |
Example
The example configuration of the repository must be adapted to the target system.
# ========================================
# Configuration file for the Nexus-console
# ========================================
# NOTE: Parameters need to be adjusted for individual system
# >> TX DEVICE (AWG) M2p.6546-x4
TxConfiguration:
# Could be extended to a list
device_path: "/dev/spcm1"
# Number of channels
max_available_channels: 4
# Device sampling rate in MHz
sampling_rate: 20
# Max. output amplitude per channel (RF, Gx, Gy, Gz) in mV
channel_max_amplitude: [200, 6000, 6000, 6000]
# Filter configuration per channel (RF, Gx, Gy, Gz)
channel_filter_type: [0, 2, 2, 2]
# Configure if RF/gradients are terminated into 50 ohm impedance (true == 50 ohm)
rf_terminated_50ohm: True
gradients_terminated_50ohm: False
# Gradient efficiency per gradient channel (Gx, Gy, Gz) in T/m/A
gradient_efficiency: [0.37e-3, 0.451e-3, 0.4e-3]
# Gradient power amplifier gain per gradient channel (Gx, Gy, Gz) in A/V
gpa_gain: [1.5, 1.5, 1.5]
# Scales RF in Hz (pulseq) to mV
rf_to_mvolt: 5.e-3
# rf_gain_lut_path: "/path/to/lut.npy"
# >> RX DEVICE (Digitizer) M2p.5933-x4
RxConfiguration:
# Could be extended to a list
device_path: "/dev/spcm0"
# Number of channels
max_available_channels: 8
# Device sampling rate in MHz
sampling_rate: 20
# Enable the receive channels, length must be equal to max_available_channels
channel_enable: [1, 0, 0, 0, 0, 0, 0, 0]
# Set the max. amplitude per channel in mV, length must be equal to max_available_channels
channel_max_amplitude: [200, 200, 200, 200, 1000, 1000, 1000, 1000]
# Configure which channels are terminated into 50 ohm impedance (true == 50 ohm), length must be equal to max_available_channels
channel_terminated_50ohm: [True, True, True, True, False, False, False, False]
SystemLimits:
# Maximum gradient amplitude in Hz/m (PyPulseq system stores this unit)
max_grad: 1703040.0 # equals 40 mT/m
# Maximum slew rate in Hz/m/s (PyPulseq system stores this unit)
max_slew: 2128800000.0 # equals 100 T/m/s
# Dead time at the beginning of RF event in s, covered by rf_delay
rf_dead_time: 20.e-6
# Time delay at the end of an RF event in s
rf_ringdown_time: 30.e-6
# Time delay at the beginning of an ADC event in s
adc_dead_time: 0.
# Raster time for block durations
block_duration_raster: 1.e-6
# Raster time for RF pulses.
rf_raster_time: 1.e-6
# Raster time for gradient waveforms.
grad_raster_time: 1.e-6
# Raster time for ADC readout.
adc_raster_time: 1.e-6